GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 7 Diagnostics/Troubleshooting
7 - 59
7-5-8-11 Tx Power Supply Test (cont’d)
Name: TxPs Control Test
Description: Transmit Power Supply Control Test (Verifies the voltage points are within specifications)
Run Time: 00:10
If this test failed:
1.) Replace the power supply
Name: TxPs Watchdog Test
Description: Transmit Power Supply GFI Watchdog Test (Tests that the power supply watchdog is
working)
Run Time: 00:07
If this test failed:
1.) Replace the power supply
Name: TxPs Alarm Test
Description: Transmit Power Supply Alarm Test (Tests the fault detection of power supply)
Run Time: 00:04
If this test failed:
1.) Replace the power supply
Name: TxPs Keep Alive Test
Description: Transmit Power Supply Keep Alive Test (Tests the communication between BEP and
power supply and fault detection)
Run Time: 00:07
If this test failed:
1.) Replace the power supply