Section 8-4 - Loading / Reloading / Upgrading the Software
4.
Press any key to continue; or to abort,
remove the disk from the drive and press
CTRL-C.
The System Software Load Instruction
window displays.
System Software load
instruction (R.4)
5.
Choose Option 1 to perform a complete
Software installation, all data will be erased.
Choose Option 2 to install Software on C
partition only. Only data on C partition will
be erased.
Choose Option 1.
The Software loader program will begin to
launch and the following warning will
appear:
You are about to reload the
entire disk. If you DO NOT want
to do this, remove the CDROM from
the CDROM drive and Press
"CTRL-C" now to exit and power
cycle your system to restart
without overwriting your disk
drive's current contents.
Press any key to continue...
The Software loader program will continue
to launch.
In R5 or later, at some point the system will
prompt to insert Disk #2. Follow the screen
instructions.
The Symantec Ghost window displays.
After any key to boot DVD Ghost
Ghost wait until next
Symantec Ghost Window
WARNING
WARNING
DO NOT interrupt the software loading at any time
.
Table 8-4
Software Load
Steps
Corresponding Graphic
Содержание LOGIQ E9
Страница 1: ...12 13 18 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL VERSION R6 Part Number 5573152 100 Revision Rev 2 ...
Страница 2: ......
Страница 11: ...GE DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL xi ZH CN KO ...
Страница 140: ...DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 50 Section 4 8 Site Log This page was intentionally left blank ...
Страница 753: ......
Страница 754: ......