Chapter 5 Components and Functions (Theory)
5 - 29
5-7-6
AC Failure Power Off Sequence
Figure 5-24 AC Failure Power Off Sequence
Main mcu detects
AC or 48V failure
TS_OK
false
48V_OK
False
Record Failure in
Flash (incl Time)
Stop
Transmitting
GFI Stops
Scanning.
Disconnect TX
from TSV.
Shutdown
BEP PS
TS_OK
Shutdown of
24V, LV
GFI or MRX
Main Power Supply
BEP
ACFAIL_N
true
ACFAIL_N
48V_OK
Shutdown
TS
Stop Scanning.
ACFAIL_N
BEP OS
shutdown of HD
and peripherals.