Section 8-8 - Casters and Brakes replacement
8-8-3-5
Front Casters installation
1.) Position the Caster so it align with the fastening screw.
2.) Install the three fixing screws (M8 x 20 mm), torque: 20.5 Nm (15.1 lbf-ft).
3.) Remove the Wooden Wedge.
4.) Roll the system off the Bevel Edged Board.
5.) To replace the other Front Caster, repeat all the steps from
8-8-3-4 "Front Casters removal" on page
but now using the Bevel Edged board and the Wooden Wedge on the other side of the
system.
6.) Install the Foot Rest Cover.
8-8-3-6
Calibration and adjustments
No calibrations or adjustments are needed after this part replacement.
8-8-3-7
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) If finished, connect cables and Probes removed earlier.
3.) Power up the system to verify that it starts as it should.
8-8-3-8
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
If all are successful, include the following debrief script: LOGIQ E9 Service Manual, Direction
5573152-100, Rev. 2. Equipment passed all required checks and is ready for use.
Table 8-128 Front Casters replacement Functional Checks
See:
Section
Functional Check
Debrief Script
LOGIQ E9 Service Manual, Direction
5573152-100, Rev. 2. Equipment passed all required
checks and is ready for use.
and
and
Содержание LOGIQ E9
Страница 1: ...12 13 18 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL VERSION R6 Part Number 5573152 100 Revision Rev 2 ...
Страница 2: ......
Страница 11: ...GE DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL xi ZH CN KO ...
Страница 140: ...DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 50 Section 4 8 Site Log This page was intentionally left blank ...
Страница 753: ......
Страница 754: ......