Section 7-5 - Common Service Desktop
7-5-3 Diagnostics Window Overview (cont’d)
11.)Select the Loop Order:
-
ABCABCABC = Runs the selected diagnostics as one group and runs each group for the
number of loops specified.
-
AAABBBCCC = Runs the selected diagnostics one at a time for the maximum loop value then
proceeds to the next diagnostic.
Some diagnostic windows that verify the tolerance of known values compile the test data and display it
in a table. You can access this data through the Details links.
12.)Click the Details icon to open the Details window.
13.)Click the See Details link to compile the Details data.
14.)The table displays the compiled detail data.
Figure 7-17 Advanced Features of Service Diagnostic Windows
Figure 7-18 Details Link on Diagnostic Window