GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 8 Replacement procedures
8 - 135
8-6-5-1
Probe Holder Insert removal
8-6-5-2
Probe Holder Insert installation
Table 8-85 Probe Holder Insert removal
Steps
Corresponding Graphic
1.
Carefully pull the flexible Probe Holder
Insert out of the lower UI frame. Large (
1
)
and small (
2
) soft, Probe Holder Inserts.
Probe Holder placement
2.
To remove the dongle, slightly lift up the end to disengage the Velcro, then pull the dongle out of
the USB port.
Table 8-86 Probe Holder Insert installation
Steps
Corresponding Graphic
1.
Install the flexible Probe Holder Insert into
the lower UI frame. Be sure the hooks (
1
)
and tabs (
2
) fit properly and the Probe
Holder fits snugly.
Probe Holder placement