GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 8 Replacement procedures
8 - 11
8-4-6-4
System Software Load
1.) Turn ON the LOGIQ E9.
2.) Turn ON all digital peripherals.
Table 8-4
Software Load
Steps
Corresponding Graphic
1.
Insert the applicable LOGIQ E9 Base
Image Load Software disk provided in the
Kit, into the DVD drive.
Power down the LOGIQ E9.
Wait until the LCDs on the Operator Panel
are switched off and the ON switch has
turned amber.
Power ON the LOGIQ E9. Some Windows
screens will be displayed before displaying
the software load warning message.
The scanner boots from the Software disk
and displays a load warning message.
System Software Load warning message
2.
Base Image consist of two disks. Insert Disk #1 (1 of 2).
3.
If “CTRL-C” to cancel is pressed:
Type “y“.
Type “exit“.
after any key is pressed
if cancelled