UUT Calibration Setup
UUT Connection Test
4
4-21
In the Panel1 plot in Figure 4-11, Panel1 shows how the phase error should look. In the
Panel1 plots in Figure 4-12, Panel1 shows approximately 100% TVE for all three current
phases. The voltage phases are also 100% TVE. This could be an indication that the
6135A forced all the outputs off because of a problem with the voltage and current
connections to the UUT. If you are not testing the current inputs and did not connect
them to the UUT, then you must set the Nominal Current to 0 in the PMU Personality
dialog. If it is not set, the 6135A forces all outputs off when it senses an overvoltage
condition on the current outputs.
gtu136.jpg
Figure 4-11. Current Phases Correct