Troubleshooting, Maintenance, and Care
Troubleshooting
8
8-7
UUT
Problem
Probable Cause
Action
This error message appears:
1. The Server PC is not
connected to the UUT.
2. Incorrect UUT IP address or
port.
3. Problem with the UUT
configuration.
4. UUT does not use
C37.118.2 Synchrophasors
Data Transmission Protocol.
Note
The UUT must be compliant
with C37.118.2 to be tested by
the Calibration System.
Complete these steps in order:
1. Make sure the UUT is on.
2. Check the Ethernet connection
to the UUT. See Figure 8-1.
3. Make sure the connection
settings in the PMU
Connection details area of the
PMU Personality Dialog box
match the UUT configuration.
Note
The UUT must be on subnet
192.92.92.92 Do not use
192.92.92.1.
This error message appears:
1. Incorrect PMU ID.
2. Problem with the UUT
configuration.
3. UUT does not use
C37.118.2 Synchrophasors
Data Transmission Protocol.
Note
The UUT must be compliant
with C37.118.2 to be tested by
the Calibration System.
Make sure the ID number in the
PMU Connection details area of
the PMU Personality Dialog box
match the UUT configuration.
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...