Automated Testing
Run an Automated Test
5
5-5
gtu107.jpg
Figure 5-2. Results File Selection
6.
Monitor the test status screen as the automated test runs (see Figure 5-3).
Note
If “Skip Current Test” is clicked, the current test is stopped and the next test in the
sequence is started. After all of the sequenced tests are complete, the tests that were
skipped can be completed. Refer to “Complete Skipped Tests” in this chapter.
If you select “Abort All Tests”, the automated test is aborted.
gtu108.jpg
Figure 5-3. Test Status Window
7.
After the test is complete, click on
Results
in the top
Navigation Menu
then click
Show Log
to open the test log (see Figure 5-5).
8.
Look in the
Status
column and make sure all tests show “Pass” or “Fail”. If there are
tests that show “Error”, “Skipped”, or “Cancelled”, use the instructions in
“Incomplete Tests” to complete the incomplete tests.
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...