6135A/PMU
Operators Manual
7-6
Table 7-1. Interactive Graph Configuration Window (cont.)
Item
Name
Description
Inputs to Plot
Use to select the which input to plot. The two selections are:
•
Voltage
– Plot the voltage test data from the source selected.
•
Current
– Plot the current test data from the source selected.
If Voltage and Current are both selected, each will open a new
window for each source selected.
Example: If both sources (actual and measured) are selected and
both inputs (voltage and current) are selected, then four windows
open: Actual Voltage, Actual Current, Measured Voltage, and
Measured Current. Refer to the examples in the “Example
Interactive Graph Configurations” section.
Phases to Plot
Use to select the phase to plot. The four selections are:
•
Phase A, B, and C
– Plot the test data for the phase selected.
•
Positive Sequence
– Plot the positive sequence for all three
phases.
X Axis
Use to select the x-axis plot. The two selections are:
•
Plot vs. Time
– Plot the test data against time.
•
Plot vs. Frequency
– Plot the test data against frequency.
If both are selected, each will open a new window for each source
and input selected. One window will show Plot vs. Time and the
other will show Plot vs. Frequency (Refer to Figure 7-2).
Y Axis: Analysis
Results
Use to select the error value to plot. The five selections are:
•
TVE (Percent)
– Plots the Total Vector Error in percentage
•
Magnitude Error
– Plot the Magnitude Error component of the
TVE.
•
Phase Error
– Plot the Phase Error component of the TVE.
•
Frequency Error
– Plot the difference (error) between the
Actual Frequency output by the System and he measured
frequency reported by the UUT.
•
ROCOF Error
- Plot the difference between the actual Rate of
Change of Frequency (ROCOF) output of the System and the
measured ROCOF of the UUT.
Each selection opens a new panel in the window.
Y Axis: Raw
Data
Use to select the raw data to plot. The four selections are:
•
Magnitude
– Plot the magnitude (amplitude) of the selected
source..
•
Phase
– Plot the phase of the selected source..
•
Frequency
– Plot the frequency of the selected source.
•
ROCOF
– Plot the Rate of Change of Frequency of the selected
source.
Each selection opens a new panel in the window.
Path
The location of the active plot settings file. Plot settings file contains
the plot selections as well as plot colors, plot styles and other
configurable plot options.
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...