General Operation Information
Turn On the Calibration System
When the Calibration System is turned on, each system will initialize and connect to the
other components in the system automatically. The front panel power switches on the
6135A Units put them into standby mode, but do not completely isolate power. The
6135A can safely be left in standby and will consume a small amount of power unless
switched off at the rear panel.
To turn on the Calibration System:
1.
Open the front panel access cover on the
Server PC
and push the
Power
button.
2.
Push the
Power
button on the rear panel of the
GPS Receiver
(if required).
3.
Push the
Power
button on the
6135A/PMU System Timing Unit
front panel.
4.
Push the
Power
button on the two
6106A (Auxiliary) L2 and L3 units
. Wait a few
minutes then push the
Power
button on the
6105A (Master) L1 unit
. If the internal
cooling fan is not heard running inside each unit, check to make sure that the rear
power switches are on. Do not operate the unit if the fan is not running.
5.
Let the Calibration System warm-up, see “Calibration System Warm-up” in this
chapter.
Turn Off the Calibration System
To turn off the Calibration System:
1.
Push the
Power
button on the
6106A (Auxiliary) units
.
2.
Push the
Power
button the two
6105A (Master) unit.
3.
Push the
Power
button on the
6135A/PMU System Timing Unit
front panel.
4.
Open the front panel access cover on the
Server PC
and push the
Power
button.
Calibration System Warm-up
To guarantee accuracy, the Calibration System must be warmed up 1 hour or more to let
the environmentally controlled components become stable.
Connect the Calibration System to a UUT
Use this section to connect the Calibration System to a UUT for test or calibration. Each
UUT has unique features that can change this connection and configuration procedure.
Before you start this procedure, read the UUT user documentation to fully understand the
limits and features of the UUT.
The basic UUT connection procedure is illustrated in Figure 4-1. This illustration lists the
procedure to connect a UUT to the Calibration System. Each topic contains references to
helpful information and procedural instructions.
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...