UUT Calibration Setup
Connect the Calibration System to a UUT
4
4-5
gtu087.jpg
Figure 4-1. UUT Connection Procedure
Prepare the UUT for Calibration
Prepare the UUT for calibration (refer to the UUTs users documentation for instructions):
1.
Read the UUT users documentation for information such as how to connect the
inputs, turn-on inputs, configure the IP address and subnet-mask, assign or find the
UUT ID number, and how to change the UUT configuration.
2.
Turn on the communication port on the UUT and set the properties to the
IEEE C37.118.2 communication standard (if required).
3.
Configure the UUT to have an IP address of
192.92.92.92
and a subnet-mask of
255.255.255.0
.
4.
Set the nominal frequency, reporting rate, and class (if configurable) to the values
necessary for the calibration procedure.
5.
Configure the bypass filters on the UUT for calibration.
Prepare the Calibration System:
1.
Turn on the Calibration System (refer to “Turn On the Calibration System” in this
Chapter).
2.
Collect all cables and connectors necessary to connect the Calibration System to the
UUT.
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...