6135A/PMU
Operators Manual
6-14
Table 6-6. Step Test Configuration Window
gtu128.emf
Item
Field Name
Range
Description
Input Frequency
(44.0 to 65.9) Hz
Set the frequency for the test.
Phase Shift
(-180
°
to 180
°
) degrees Set the phase offset at the start of the test. All test
signals are specified to be cosine waves.
Phase Step Size
(-180
°
to 180
°
) degrees Set the amount of phase in degrees applied at the
step.
Amplitude Step
Size
(-100 to +200) %
Set the amount of voltage or current magnitude in
percentage of nominal voltage applied at the step.
Seconds Until
Step
(0 to 255) Seconds
Set the second that the step is accomplished.
Number of Steps 1 to 40
Set the how many step test iterations to run.