Calibration Procedure Information
The UUT calibration procedure illustrated in Figure 3-2 is a remotely operated procedure
that is controlled by the PMUCal Software. To certify a UUT, the PMUCal Software is
configured to run preprogrammed automated certification tests (the certification tests). A
certification test is an automated test file that contains a noneditable list of sequenced
tests (refer to “Automated Testing” in this chapter). A certification test contains more
than 1,000 individual sequenced tests derived from the standards. Each certification test
is configured for a nominal UUT frequency, a UUT reporting rate, and a UUT class.
When a test is run, the PMUCal Software sends data to the Server PC to configure the
test. The Server PC uses the data to configure and initialize the Calibration System. At
the set test start time, the Server PC starts the test and actively controls all of the
Calibration System outputs.
While the test is in progress, the Server PC records the reference data from the 6135A
System and the measured data from the UUT. The reference and measured data is sent to
the Client PC and the maximum test values are saved to the active test results file (see
“Standard Compliance Tests and Test Files” in this chapter).
When the test is complete, the Calibration Software updates the active test results file
which can viewed on the Results screen or can be exported to Microsoft Excel file. If the
test done was a certification test and all tests passed, a formal certification report can be
generated.
For more information on Automated Testing and Interactive Testing, refer to “Testing
Methods” in this chapter.
gtu057.jpg
Figure 3-2. How the Calibration System Calibrates a UUT
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...