6135A/PMU
Operators Manual
6-12
Table 6-5. Modulation Test Configuration Window
gtu127.emf
Item
Field Name
Range
Description
Input Frequency
(44.0 to 65.9) Hz
Set the frequency for the test.
Phase Shift
(-180
°
to 180
°
) degrees Set the phase offset at the start of the test. All test
signals are specified to be cosine waves.
Phase Modulation
Frequency
(0.1 to 12) Hz
Set the frequency of the phase modulation.
Phase Modulation
Index
(0, 1, and 10)
%
Set the amplitude of phase shift.
Amplitude
Modulation
Frequency
(0.1 to 12) Hz
Set the frequency of the amplitude modulation.
Amplitude
Modulation Index
(0 to 30) %
Set the amount of voltage or current amplitude in
percentage to be applied to the frequency.
Settling Time
(0 to 255) reports
Set the number of UUT reports to not include in the
analysis or the reports at the start and end of the test
duration.
Note
The duration must be at least 1 second longer than
the settling time.
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...