6135A/PMU
Operators Manual
5-6
Incomplete Tests
After the automated test is complete, use the results log to complete incomplete tests. An
incomplete test is defined as a test that has a status of “Cancelled”, “Skipped”, or “Error”
on the log.
To complete skipped automated tests:
Note
Skip to step 5 if you already have the applicable Results file loaded.
1.
Prepare the UUT for test and calibration (use the instructions in “Prepare the UUT
for Calibration” in Chapter 5).
2.
Click
Results
in the top
Navigation Menu
.
3.
Click
Load/New
to load the PMU Personality of the PMU. The results file selection
menu opens.
4.
Use the selections to load the results file that contains the results for the automated
test that was done on the UUT.
Note
The results files are marked with time stamps to identify when the results
file was made (refer to “Result Files Naming Convention” in Chapter 3). In
addition to the result file time stamp, Date Modified column in the load
dialog box can be used to find a recently modified results file (see
Figure 5-4).
gtu110.jpg
Figure 5-4. Load Results File Dialog Box
5.
Click
Results
in the top
Navigation Menu
.
6.
Click
Show Results
to view the skipped or incomplete tests.
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...