6135A/PMU
Operators Manual
4-16
Table 4-3. PMU Phase Order Screen (cont.)
Item
Name
Description
Connect to PMU
and Get PMU
Config
Connects to the PMU and returns the configuration.
Save
Use to save the current UUT personality profile.
Load
Use to select and load a saved UUT personality profile.
UUT Connection Test
After a connection to the UUT is made, run a default steady state test to make sure all
UUT connections are made correctly as follows:
1.
Click
Test
in the top
Navigation Menu
.
2.
In
Interactive Testing
, click
Customize and Run Single Tests
.
3.
Click on
Load Defaults
.
4.
Click on
Run Test
to run a default steady state test on the connected UUT as shown
in Figure 4-6.
Note
The Input Frequency should match the Nominal Frequency set in the PMU
Personality.
gtu0122.png
Figure 4-6. Test Menu