6135A/PMU
Operators Manual
2-4
Install the Calibration System
This section supplies information on how to install the Calibration System. Use this
section for the initial installation and also when the Calibration System comes back from
annual adjustment or repair.
Pre-Installation Preparation
Placement preparation is important for a successful Calibration System installation. The
Calibration System needs to have access to three independent power outlets, a roof
mounted GPS antenna, and a network connection for remote operation. Use the
considerations in this section to help determine where to put the Calibration System.
Electrical Considerations:
•
The maximum total power requirement for the Product is approximately 3500 VA.
Three power cords are provided to share the load of the Product between three
standard power outlets. Refer to the “General Specifications” section of this manual
for input power requirements.
•
The Calibration System must have a grounded power supply.
•
The central earth connection for the equipment in the system is located in the bottom
rear of the cabinet which is connected to earth via a connection on the Power
Distribution Unit (PDU) and the power cords.
•
The equipment in the cabinet and the PMU being tested are powered by the PDU in
the rear of the cabinet. The equipment within the cabinet and the PMU being tested
are protected by resettable breakers in the PDU.
Environmental Considerations
:
Refer to the “General Specifications” section of this manual for environmental condition
limits.
Cooling Considerations:
Product must be located at least 4 inches (100 mm) away from any objects that could
restrict air flow. Make sure that the intake air to the Product is within the operating
temperature limits in the environmental specification.
W
Caution
To prevent damage to the Product from overheating, make
sure the space around the Product is not less than 4 inches
(100 mm).
Note
Do not put or store objects on top of the Calibration System as the top
of the server rack can become very warm.
Other Considerations
:
The Calibration System needs to connect to a separate PC to operate. Install the
Calibration System in an area where there is direct access to a PC or a network
connection.
For first time installations:
1.
Install the Symmetricom roof mounted GPS Antenna. For instructions, refer to the
Microsemi SyncServer S650 Users Guide.
2.
Connect the GPS Antenna cable from the GPS Antenna to where the Calibration
System is to be located. For instructions, refer to the Microsemi SyncServer S650
Users Guide.
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...