6135A/PMU
Operators Manual
4-20
In Panel2 plot in Figure 4-10, Panel2 shows a Magnitude error of 70 volts on the phase b
voltage channel. The TVE for phase B is 100%. These inaccuracies indicate that the
voltage input is unconnected.
gtu137.jpg
Figure 4-10. Problem - Voltage Phase B is Not Connected