Software Operation
About this Chapter
3
3-3
About this Chapter
This chapter supplies information on the PMUCal Software supplied with the Calibration
System. The beginning of the chapter provides information on the UUT calibration
procedure that is controlled by the PMUCal Software. This information is followed by
functionality and feature information to help understand the different testing methods and
modes of operation available. For an alphabetical list of terms and abbreviations used in
this chapter and manual, see “The Glossary” in Appendix A of this manual.
UUT Calibration Procedure
Overview
UUT calibration is a performance verification procedure to make sure that a UUT is in
accordance with the performance requirements specified in
IEEE C37.118.1a-2014,
Synchrophasor Measurements for Power Systems Specification
. The Product compares
phasors from the UUT with the reference voltage and current phasors it generates.
Note
The word “reference” or “actual” in this manual refers to the calibrated
source output signals from the 6135A System to the UUT. The word
“measured” refers to the measured signals from the UUT outputs.
The basic UUT calibration procedure is illustrated in Figure 3-1. This illustration shows
the procedure to calibrate or test a UUT. Each topic references to helpful information and
procedural instructions in this manual.
Calibration Procedure Start
Prepare for UU T Calibration. Read the UU T user documenation and configure the UU
T for testing.
For UU T preparation procedures, refer to “
Prepare the UUT for Calibration
” in Chapter 4.
Testing Method Selection
Choose from certification, automated, and general testing methods. For information,
refer to “ Testing Methods” in Chapter 3.
Run the Test
Run the test on the UU T. For test procedures, refer to “
Automated Testing
” and
“
General Testing
” in Chapter 5 and 6.
View the Results
The test results can be viewed in the results screen or exported to a Microsoft
Excel Spreadsheet. If the test was an interactive test, the results can be viewed on
an interactive graph. For information, refer to “
Test Results
” in Chapter 7.
Create a Report
A test report can be made from the test results. If the was a certification test,
a formal calibration report can be made. For procedures on how to make a
test report, refer to “
Test Reports
” in Chapter 7.
Calibration Procedure End
After the test is complete, you can do the procedure again with a
different UU T configuration or testing method to troubleshoot the
UUT if a test failed.
gtu083.emf
Figure 3-1. Calibration Procedure with References
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...