6135A/PMU
Operators Manual
6-8
Steady State Test Parameters
A Steady State test is where the magnitude, and frequency are constant for the duration of
the test (see “Standard Compliance Tests and Test Files” in Chapter 3 for more
information). See Table 6-3 for a detailed description of each parameter on the Steady
State test window.
To configure and run a Steady State test (refer to Table 6-3):
1.
Set the
General Test Parameters
. (refer to “General Test Parameters” in this
section).
2.
Click
Steady State Tests
.
3.
Set the
Steady State Test Parameters
as follows:
a.
Set the
Input Frequency
.
b.
Set the
Phase Shift
.
c.
Set the
Voltage Magnitude Index
.
d.
Set the
Current Magnitude Index
.
e.
Set the
Harmonics Number, Phase
, and
Index
.
f.
Set the
Interharmonic Frequency
,
Phase
, and
Index
.
4.
Click
Save
to save the test (if desired).
5.
Run the test. Refer to the instructions in “Run a Custom Interactive Test” in this
chapter.
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...