Installation Instructions
About this Chapter
2
2-3
About this Chapter
This chapter contains information on how to safely install and configure the Calibration
System. To become familiar with the Calibration System, read Chapter 1before
installation begins. For an alphabetical list of terms and abbreviations used in this chapter
and manual, see “The Glossary” in Appendix A of this manual.
Unpack and Inspect
The Calibration System is shipped in a wood shipment crate. Use the shipment crate to
transport or ship the calibrator to Fluke Calibration for adjustment or repair.
Note
After the Calibration System is removed from the crate, store the crate in a
dry, safe area for future use.
To unpack, put the crate in a flat, level area with sufficient room to lower the crate ramp
and maneuver the Calibration System. Visually examine the crate for shipment damage
indications.
Note
If the crate has visible shipment damage, do not open the crate. Photograph
the damage and contact the shipment company and Fluke (refer to
“Contact Fluke Calibration” in Chapter 1).
To unpack the Calibration System:
1.
Unfasten the door to the shipment crate.
2.
Remove the door from the shipment crate.
3.
Set the door upside down on the ground to use as a ramp.
4.
Remove the wooden braces.
5.
Inspect system cabinet for shipment damage and ensure that the cabinet door is
secure.
W
Warning
The Calibration System weighs approximately 170 kg (375 lb).
To prevent personal injury or damage to the Calibration System
use two or more people to move the Calibration System out of
the crate and slowly down the ramp.
Note
The tight packaging inside the shipment crate can make it hard to move
the Calibration System out of the crate. The bottom frame on the front
of the Calibration System is a good location to hold and pull.
6.
Visually examine the Calibration System for damage.
7.
Remove and open the three accessory boxes found in the top shelf inside the
shipment crate.
8.
Examine to make sure all the standard accessories in Table 1-1 are in your
accessories kit.
9.
Reassemble the shipment crate.
10.
Store the shipment crate in a safe area for future use.
Содержание 6135A/PMU
Страница 28: ...6135A PMU Operators Manual 2 2...
Страница 44: ...6135A PMU Operators Manual 3 2...
Страница 70: ...6135A PMU Operators Manual 3 28...
Страница 72: ...6135A PMU Operators Manual 4 2...
Страница 92: ...6135A PMU Operators Manual 4 22 gtu135 jpg Figure 4 12 Problem All Outputs Forced Off...
Страница 94: ...6135A PMU Operators Manual 5 2...
Страница 102: ...6135A PMU Operators Manual 5 10...
Страница 104: ...6135A PMU Operators Manual 6 2...
Страница 120: ...6135A PMU Operators Manual 7 2...
Страница 138: ...6135A PMU Operators Manual 7 20 gtu120 emf Figure 7 18 Example Report...
Страница 140: ...6135A PMU Operators Manual 8 2...
Страница 150: ...6135A PMU Operators Manual 9 2...
Страница 160: ...6135A PMU Operators Manual A 4...