18
Low Level Disinfection (optional)
1. Disinfect the device in a disinfecting solution that has one of the following
active Ingredients:
• ≥ 2.4% glutaraldehyde
4
with a minimum soaking time
of 45 minutes at 25 °C
or
• ≥ 3.4% glutaraldehyde
5
with a minimum soaking time
of 20 minutes at 25 °C
or
• ≥ 0.55% ortho-phthalaldehyde
6
with a minimum soaking time
of 12 minutes at 25 °C.
2. Prepare the disinfecting solution according to the manufacturer’s
instructions.
3. Per manufacturer’s recommendations, immerse the device, filling all
lumens, in the disinfecting solution for the required time at the appropriate
temperature.
4. Thoroughly rinse and flush all parts and lumens with running,
demineralized water to remove the disinfectant.
5. Dry all parts with a lint-free towel immediately after rinsing.
4
CIDEX Activated® is validated for disinfection efficacy.
5
CIDEX Plus® is validated for disinfection efficacy.
6
CIDEX® OPA is validated for disinfection efficacy.
Drying
• For automated drying, use the drying cycle provided with the washer/
disinfector.
• For manual drying, use a lint-free cloth.
• Dry any lumens with compressed air.
Maintenance, Inspection, and Testing
• Inspect the device on a continual basis. If a problem is observed
or suspected, the device should be returned for repair.
• Inspect all components for cleanliness. If fluid or tissue buildup
is present, repeat the above cleaning and disinfection procedures.
• Inspect the camera cable for cuts and breaks. Return any damaged
urology camera to Stryker for service.
Packaging
N/A
Summary of Contents for 1288
Page 1: ...User Guide 1288 HD Urology Camera Head 1288310130 ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 17: ...13 Adjusting the Focus Slide the focusing knob to the left or right to adjust the focus ...
Page 26: ......
Page 28: ......
Page 48: ......
Page 50: ......
Page 72: ......
Page 94: ......
Page 116: ......
Page 118: ......
Page 138: ......
Page 140: ......
Page 151: ...147 De focus aanpassen Schuif de focusknop naar links of rechts om de focus aan te passen ...
Page 162: ......
Page 173: ...169 Justering af fokus Skub fokuseringsknappen til venstre eller højre for at justere fokus ...
Page 182: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 217: ...213 Justering av fokus Skyv fokuseringsknappen til venstre eller høyre for å justere fokus ...
Page 226: ......
Page 228: ......
Page 239: ...235 Ställa in fokus Vrid fokuseringsratten åt vänster eller höger för att justera fokus ...
Page 248: ......
Page 250: ......
Page 261: ...257 Regulacja ostrości Przekręć pokrętło ogniskujące w lewo lub prawo aby dopasować ogniskową ...
Page 272: ......
Page 294: ......
Page 301: ...297 将内窥镜插入内窥镜体夹子中 向相反的方向转动内窥镜体闸以固定内窥镜 4 将光缆从光源连接到内窥镜上的光源接线柱上 ...
Page 305: ...301 调焦 将调焦旋钮滑向左边或右边以调整焦距 ...
Page 314: ......
Page 316: ......
Page 327: ...323 焦点の調整 フォーカスノブを左または右にスライドさせて フォーカスを調整 します ...
Page 336: ......
Page 338: ......
Page 349: ...345 초점 조정 Adjusting the Focus 초점 손잡이를 왼쪽 또는 오른쪽으로 밀어 초점을 조정하십시오 ...
Page 358: ......
Page 359: ......