17
Cleaning: Automated
1. Brush
• Using a syringe, inject 50mL of the enzymatic detergent (from the
“Preparation for Cleaning” section) into any lumen and mated surface
a minimum of one time.
• Brush from both ends of any lumens a minimum of 1 time, using an
appropriate bottle brush.
2. Rinse
• Rinse the device with RO/DI water at ambient temperature until there
is no visible detergent residue. Continue to rinse for a minimum of
30 seconds after all detergent residue has been removed.
• Place the device in the washer on an incline to facilitate drainage.
3. Automated wash
• Program the washer using the following parameters:
Phase
Recirculation
Time
Water
Temperature
Detergent Type and
Concentration
(if applicable)
Pre Wash
2 minutes
Cold tap
water
N/A
Enzyme
Wash
2 minutes
Hot tap water
Enzymatic Detergent
2
Wash 1
2 minutes
Set point
(66°C)
Non-enzymatic Detergent
3
Rinse 1
2 minutes
Hot tap water
N/A
Dry Phase
7 minutes
115°C
N/A
• Upon completion of the draining phase after Rinse 1, stop the cycle and
open the washer door.
• Remove the device from the washer during the thermal phase and place
the device back into the washer for the dry phase.
• If necessary, use pressurized air to aid in drying. Visually inspect each
device for cleanliness.
2
ENZOL® Enzymatic Detergent is validated for cleaning efficacy.
3
Renu-Klenz® is validated for cleaning efficacy.
Summary of Contents for 1288
Page 1: ...User Guide 1288 HD Urology Camera Head 1288310130 ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 17: ...13 Adjusting the Focus Slide the focusing knob to the left or right to adjust the focus ...
Page 26: ......
Page 28: ......
Page 48: ......
Page 50: ......
Page 72: ......
Page 94: ......
Page 116: ......
Page 118: ......
Page 138: ......
Page 140: ......
Page 151: ...147 De focus aanpassen Schuif de focusknop naar links of rechts om de focus aan te passen ...
Page 162: ......
Page 173: ...169 Justering af fokus Skub fokuseringsknappen til venstre eller højre for at justere fokus ...
Page 182: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 217: ...213 Justering av fokus Skyv fokuseringsknappen til venstre eller høyre for å justere fokus ...
Page 226: ......
Page 228: ......
Page 239: ...235 Ställa in fokus Vrid fokuseringsratten åt vänster eller höger för att justera fokus ...
Page 248: ......
Page 250: ......
Page 261: ...257 Regulacja ostrości Przekręć pokrętło ogniskujące w lewo lub prawo aby dopasować ogniskową ...
Page 272: ......
Page 294: ......
Page 301: ...297 将内窥镜插入内窥镜体夹子中 向相反的方向转动内窥镜体闸以固定内窥镜 4 将光缆从光源连接到内窥镜上的光源接线柱上 ...
Page 305: ...301 调焦 将调焦旋钮滑向左边或右边以调整焦距 ...
Page 314: ......
Page 316: ......
Page 327: ...323 焦点の調整 フォーカスノブを左または右にスライドさせて フォーカスを調整 します ...
Page 336: ......
Page 338: ......
Page 349: ...345 초점 조정 Adjusting the Focus 초점 손잡이를 왼쪽 또는 오른쪽으로 밀어 초점을 조정하십시오 ...
Page 358: ......
Page 359: ......