14
Cleaning, Reprocessing, and
Maintenance
Reprocessing the Urology Camera Head
These reprocessing instructions are provided in accordance with ISO 17664,
AAMI TIR12, AAMI ST79, and AAMI ST81. While they have been validated
by Stryker as being capable of preparing the device for re-use, it remains the
responsibility of the processor to ensure that the reprocessing as actually
performed, using equipment, materials, and personnel in the reprocessing
facility, achieves the desired result. This normally requires validation and
routine monitoring of the process. Stryker recommends users observe these
standards when reprocessing medical devices.
Warnings
• This device must be cleaned and sterilized prior to the first use and after
every subsequent use.
• Use only the sterilization cycles outlined in this document. Using
unspecified sterilization cycles may damage the device or result in
incomplete sterilization.
• Separate the urology camera head and scope prior to cleaning,
disinfection, or sterilization.
• Wear appropriate protective equipment: gloves, eye protection, etc.
Cautions
• Always install the soaking cap prior to processing the urology camera.
Failure to properly tighten the soaking cap will corrode the connector
pins and void the warranty.
• Inspect the camera cable for cuts and breaks before soaking in any fluid.
Return any damaged camera to Stryker for service.
• Do not use brushes or pads with metal or abrasive tips during manual
cleaning, as permanent scoring or damage could result.
• To minimize galvanic corrosion, avoid soaking dissimilar metals in close
proximity.
• Never soak the camera in the same tray with sharp instruments.
• Repeated sterilization via Ethylene Oxide may result in degradation
in image quality.
Summary of Contents for 1288
Page 1: ...User Guide 1288 HD Urology Camera Head 1288310130 ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 17: ...13 Adjusting the Focus Slide the focusing knob to the left or right to adjust the focus ...
Page 26: ......
Page 28: ......
Page 48: ......
Page 50: ......
Page 72: ......
Page 94: ......
Page 116: ......
Page 118: ......
Page 138: ......
Page 140: ......
Page 151: ...147 De focus aanpassen Schuif de focusknop naar links of rechts om de focus aan te passen ...
Page 162: ......
Page 173: ...169 Justering af fokus Skub fokuseringsknappen til venstre eller højre for at justere fokus ...
Page 182: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 217: ...213 Justering av fokus Skyv fokuseringsknappen til venstre eller høyre for å justere fokus ...
Page 226: ......
Page 228: ......
Page 239: ...235 Ställa in fokus Vrid fokuseringsratten åt vänster eller höger för att justera fokus ...
Page 248: ......
Page 250: ......
Page 261: ...257 Regulacja ostrości Przekręć pokrętło ogniskujące w lewo lub prawo aby dopasować ogniskową ...
Page 272: ......
Page 294: ......
Page 301: ...297 将内窥镜插入内窥镜体夹子中 向相反的方向转动内窥镜体闸以固定内窥镜 4 将光缆从光源连接到内窥镜上的光源接线柱上 ...
Page 305: ...301 调焦 将调焦旋钮滑向左边或右边以调整焦距 ...
Page 314: ......
Page 316: ......
Page 327: ...323 焦点の調整 フォーカスノブを左または右にスライドさせて フォーカスを調整 します ...
Page 336: ......
Page 338: ......
Page 349: ...345 초점 조정 Adjusting the Focus 초점 손잡이를 왼쪽 또는 오른쪽으로 밀어 초점을 조정하십시오 ...
Page 358: ......
Page 359: ......