16
Cleaning: Manual
1. Brush
• Prepare a fresh solution of enzymatic detergent according to the
manufacturer’s recommendations (one ounce per gallon of tap water
at 35 - 40°C)
2
.
• Thoroughly brush the exterior of the device with a soft-bristled brush,
focusing on any mated or rough surfaces.
• Using a syringe, inject any lumen or mated surface a minimum of
5 times with 50mL of the detergent.
• Brush any lumens a minimum of 5 times from each end, using an
appropriate bottle brush.
• Brush any movable parts in all extreme positions.
2. Rinse
• Rinse the device with reverse osmosis/de-ionized (RO/DI) water at
ambient temperature until all detergent residue is removed. Flush any
lumens or mated surfaces a minimum of 5 times. Once all detergent
residue is removed, continue to rinse for a minimum of 30 seconds.
• Drain excess water from the device and dry it using a clean cloth or
pressurized air.
• Visually inspect the device for cleanliness, paying close attention to
hard-to-reach areas. If visible soil remains, repeat steps 1 and 2.
3. Soak
• Prepare a non-enzymatic detergent according to the manufacturer’s
recommendations (0.25 ounces per gallon of tap water at 35 - 40°C)
3
.
• Fully immerse the device and use a syringe to inject any lumens and
mated surfaces with 50mL of the detergent.
• Soak the device for a minimum of 15 minutes.
4. Brush
• Thoroughly brush the exterior of the device using a soft-bristled brush.
• Using a syringe, inject 50mL of the detergent into any cannulae, lumens,
or mated surfaces a minimum of 5 times.
• Brush any lumens a minimum of 5 times from each end, using an
appropriate bottle brush.
• Actuate the device, brushing around any movable parts in all extreme
positions.
5. Rinse
• Thoroughly rinse the device with RO/DI water until all detergent
residue is removed. Flush any lumens or crevices a minimum of 5 times.
Once all detergent residue is removed, continue to rinse for a minimum
of 30 seconds.
• Drain the excess water from the device and dry it using a clean cloth or
pressurized air.
Summary of Contents for 1288
Page 1: ...User Guide 1288 HD Urology Camera Head 1288310130 ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 17: ...13 Adjusting the Focus Slide the focusing knob to the left or right to adjust the focus ...
Page 26: ......
Page 28: ......
Page 48: ......
Page 50: ......
Page 72: ......
Page 94: ......
Page 116: ......
Page 118: ......
Page 138: ......
Page 140: ......
Page 151: ...147 De focus aanpassen Schuif de focusknop naar links of rechts om de focus aan te passen ...
Page 162: ......
Page 173: ...169 Justering af fokus Skub fokuseringsknappen til venstre eller højre for at justere fokus ...
Page 182: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 217: ...213 Justering av fokus Skyv fokuseringsknappen til venstre eller høyre for å justere fokus ...
Page 226: ......
Page 228: ......
Page 239: ...235 Ställa in fokus Vrid fokuseringsratten åt vänster eller höger för att justera fokus ...
Page 248: ......
Page 250: ......
Page 261: ...257 Regulacja ostrości Przekręć pokrętło ogniskujące w lewo lub prawo aby dopasować ogniskową ...
Page 272: ......
Page 294: ......
Page 301: ...297 将内窥镜插入内窥镜体夹子中 向相反的方向转动内窥镜体闸以固定内窥镜 4 将光缆从光源连接到内窥镜上的光源接线柱上 ...
Page 305: ...301 调焦 将调焦旋钮滑向左边或右边以调整焦距 ...
Page 314: ......
Page 316: ......
Page 327: ...323 焦点の調整 フォーカスノブを左または右にスライドさせて フォーカスを調整 します ...
Page 336: ......
Page 338: ......
Page 349: ...345 초점 조정 Adjusting the Focus 초점 손잡이를 왼쪽 또는 오른쪽으로 밀어 초점을 조정하십시오 ...
Page 358: ......
Page 359: ......