4-4
Rear Panel
SRS Residual Gas Analyzer
Rear Panel
The rear panel of the standard ECU Box has two connectors, two locking knobs, a
cooling fan, and eight LED’s.
Units with the optional, built-in power module (Option 02) also have a fused Power
Entry Module with a built-in power switch.
RS232/DCE/28.8k
+24VDC @ 2.5A
Power
Filament
Degas
ElecMult
RS232
Error
Leak
Burnt
Lock
Lock
Input Voltage
90 - 264 VAC
Input Frequency
47 - 63 Hz
Input voltage is set automatically. Use a line
voltage and frequency within ranges at right.
Chassis
Ground
Locking knob
LEDs
Cooling fan
RS232 connector
Power connector
Built in power module*
*option 02
Figure 3 ECU Rear Panel
[email protected] Connector
Use this connector to provide external power to the RGA. The ECU electronics must be
powered with a 24 V (+/- 2V) DC power supply @2.5Amps. The power supply must
have a cable with a 9 pin, type D, female connector on the free end, wired as described in
the “ECU- 24VDC power connector” diagram shown below.
Summary of Contents for RGA100
Page 4: ...SRS Residual Gas Analyzer iv...
Page 18: ...xviii Command List SRS Residual Gas Analyzer...
Page 46: ...2 14 Residual Gas Analysis Basics SRS Residual Gas Analyzer...
Page 66: ......
Page 78: ...4 12 Mass Filter Power supply SRS Residual Gas Analyzer...
Page 104: ......
Page 107: ...Programming the RGA Head 6 3 SRS Residual Gas Analyzer Error Byte Definitions 6 69...
Page 216: ...8 26 Quadrupole filter cleaning SRS Residual Gas Analyzer...
Page 246: ...11 2 SRS Residual Gas Analyzer...
Page 247: ......
Page 248: ......
Page 268: ...Appendix B SRS Residual Gas Analyzer 7...
Page 312: ...Appendix D SRS Residual Gas Analyzer 27...