Error Byte Definitions 6-69
SRS Residual Gas Analyzer
Error Byte Definitions
The
Error Bytes
described in this section store the results of the firmware-driven checks
built into the RGA Head. Use the Error Reporting commands to query the value of the
bytes.
Important
: No errors are present as long as all bits in the Error Bytes are cleared.
The
RGA Windows
software supports all the Error Reporting commands and reports the
errors detected based on their Error Codes.
The RGA Troubleshooting chapter identifies the different problems based on their Error
Codes and suggests possible causes and solutions.
Bit
Internal Check
(a)
ERROR Byte
(c)
Query
(d)
7
not used
N.A.
N.A.
6
24V External P/S
(b)
PS_ERR
EP?
5
Electrometer
(b)
DET_ERR
ED?
4
Quadrupole Mass Filter
(b)
QMF_ERR
EQ?
3
Electron Multiplier
CEM_ERR
EM?
2
not used
N.A.
N.A.
1
Filament
FIL_ERR
EF?
0
Communications
RS232_ERR
EC?
(a) RGA component checked. Each check involves several tests on the component.
(b) Checks automatically performed upon a power-on-reset
(c) Error bytes that store the results of the tests for each type of internal check.
(d) Error Reporting command that queries the error byte.
STATUS Error Byte
: General Status byte. Each bit corresponds to a different type of
internal check.
Summary of Contents for RGA100
Page 4: ...SRS Residual Gas Analyzer iv...
Page 18: ...xviii Command List SRS Residual Gas Analyzer...
Page 46: ...2 14 Residual Gas Analysis Basics SRS Residual Gas Analyzer...
Page 66: ......
Page 78: ...4 12 Mass Filter Power supply SRS Residual Gas Analyzer...
Page 104: ......
Page 107: ...Programming the RGA Head 6 3 SRS Residual Gas Analyzer Error Byte Definitions 6 69...
Page 216: ...8 26 Quadrupole filter cleaning SRS Residual Gas Analyzer...
Page 246: ...11 2 SRS Residual Gas Analyzer...
Page 247: ......
Page 248: ......
Page 268: ...Appendix B SRS Residual Gas Analyzer 7...
Page 312: ...Appendix D SRS Residual Gas Analyzer 27...