Troubleshooting the RGA communications 6-11
SRS Residual Gas Analyzer
Troubleshooting the RGA communications
The RGA’s communication interface includes a complete set of tools for troubleshooting
communications between a host computer and the spectrometer’s head during
programming:
1. Visual Clue
The first clue of a communication error can be obtained watching the Error LED while
communicating with the head.
The ERROR_LED is flashed on/off two times every time a communication error is
detected.
2. Error Queries
The presence of a communication error can also be detected queering the contents of the
STATUS byte in the RGA head. The query command
ER?
returns the value of the
STATUS Byte for analysis, and a communication error is present if Bit 0 of STATUS is
found set. The specific problem is determined querying the bits of RS232_ERR byte
using the
EC?
which clears the RS232_ERR byte and Bit 0 of STATUS at the end of its
execution to provide a clean error-reporting slate.
Communication errors accumulate in the RS232_ERR byte as they occur and the byte
can only be cleared by reading it with the EC? command.
Note:
Please refer to the “Error Byte Definitions” section of this chapter for a complete
listing of the error and status bytes of the RGA. Also consult the “Error Reporting
Commands” list for details on the
EC
and
ER
commands.
Summary of Contents for RGA100
Page 4: ...SRS Residual Gas Analyzer iv...
Page 18: ...xviii Command List SRS Residual Gas Analyzer...
Page 46: ...2 14 Residual Gas Analysis Basics SRS Residual Gas Analyzer...
Page 66: ......
Page 78: ...4 12 Mass Filter Power supply SRS Residual Gas Analyzer...
Page 104: ......
Page 107: ...Programming the RGA Head 6 3 SRS Residual Gas Analyzer Error Byte Definitions 6 69...
Page 216: ...8 26 Quadrupole filter cleaning SRS Residual Gas Analyzer...
Page 246: ...11 2 SRS Residual Gas Analyzer...
Page 247: ......
Page 248: ......
Page 268: ...Appendix B SRS Residual Gas Analyzer 7...
Page 312: ...Appendix D SRS Residual Gas Analyzer 27...