7-10 Peak Tuning Procedure
SRS Residual Gas Analyzer
x
For large temperature changes:
The sensitivity of the RF power supply to its
controlling voltages might be affected or, more fundamentally, the relationship
between mass and RF levels in the filter might change ( for example, if the QMF
changes its physical dimensions). In this case a Peak Tuning procedure will be
necessary to reestablish the mass axis scale.