10-16 Description of Schematics
SRS Residual Gas Analyzer
multiplier output is compared to the set level (HV_SET) by an error amplifier (U800A)
which controls the voltage on the center tap of the primary via Q800. The primary current
in the HV inverter is sensed by R814, a 2.2
:
resistor. If the primary drive exceeds 230
mA, the output of the difference amplifier U800B will go positive, reducing down the
primary drive level.
The HV power supply piggy-backs off the vertical interconnect PCB in between the top
and bottom PCBs. The HV power supply will only be installed in units which are ordered
with the channel electron multiplier option.
Summary of Contents for RGA100
Page 4: ...SRS Residual Gas Analyzer iv...
Page 18: ...xviii Command List SRS Residual Gas Analyzer...
Page 46: ...2 14 Residual Gas Analysis Basics SRS Residual Gas Analyzer...
Page 66: ......
Page 78: ...4 12 Mass Filter Power supply SRS Residual Gas Analyzer...
Page 104: ......
Page 107: ...Programming the RGA Head 6 3 SRS Residual Gas Analyzer Error Byte Definitions 6 69...
Page 216: ...8 26 Quadrupole filter cleaning SRS Residual Gas Analyzer...
Page 246: ...11 2 SRS Residual Gas Analyzer...
Page 247: ......
Page 248: ......
Page 268: ...Appendix B SRS Residual Gas Analyzer 7...
Page 312: ...Appendix D SRS Residual Gas Analyzer 27...