1-12 Installation
SRS Residual Gas Analyzer
2. Insert the RGA Windows software CD into the CD-drive of the computer.
3. Run the Setup program.
4. The software is automatically and completely installed by the RGA Setup Wizard.
Read and follow all instructions.
5. An “RGA” program group with the “RGA 3.0” icon is automatically created at the
end of the installation process.
6. Take a moment at this time to read the RGA Windows Software Chapter of this
manual.
7. Connect any third party USB-to-Serial or Ethernet-to-serial adapter to the computer
and install all necessary software drivers as indicated by its manufacturer before
attempting to run the RGA Windows software.
Turning on the RGA
1. Power up the RGA Head:
Standard RGA heads are powered up by turning on the external 24V power supply.
Units with a built in power module (Option 02) have a power switch on the back
panel of the ECU box.
2. When power is applied to the ECU, a firmware routine automatically checks the
external voltage level and turns on the green Power LED if the voltage is within the
acceptable range of 24 +/-2 V DC. The SRS RGA Head is now ready to
communicate with the computer.
Summary of Contents for RGA100
Page 4: ...SRS Residual Gas Analyzer iv...
Page 18: ...xviii Command List SRS Residual Gas Analyzer...
Page 46: ...2 14 Residual Gas Analysis Basics SRS Residual Gas Analyzer...
Page 66: ......
Page 78: ...4 12 Mass Filter Power supply SRS Residual Gas Analyzer...
Page 104: ......
Page 107: ...Programming the RGA Head 6 3 SRS Residual Gas Analyzer Error Byte Definitions 6 69...
Page 216: ...8 26 Quadrupole filter cleaning SRS Residual Gas Analyzer...
Page 246: ...11 2 SRS Residual Gas Analyzer...
Page 247: ......
Page 248: ......
Page 268: ...Appendix B SRS Residual Gas Analyzer 7...
Page 312: ...Appendix D SRS Residual Gas Analyzer 27...