6-72 Error Byte Definitions
SRS Residual Gas Analyzer
Bit
Description
Error Code
7
No filament detected.
FL7
6
Unable to set the requested emission current.
FL6
5
Vacuum Chamber pressure too high.
FL5
4
Not used
3
Not used
2
Not used
1
Not used
0
Single filament operation.
FL0
FIL_ERR Error Byte
: Filament Error Byte.
Bit
Description
7
Not used
6
Parameter conflict
5
Jumper protection violation
4
Transmit buffer overwrite
3
OVERWRITE in receiving
2
Command-too-long.
1
Bad Parameter received
0
Bad command received
RS232_ERR Error Byte