2 Programming the Test Set
STATus Subsystem
Remote Command
:STATus:QUEStionable:POWer:PTRansition <integer>
:STATus:QUEStionable:POWer:PTRansition?>
Example
STAT:QUES:POW:PTR 32 50 MHz Input Power became too high for Cal will be reported to the
Power Summary of the Status Questionable register.
Preset
32767
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Temperature Register
"Questionable Temperature Condition" on page 82
"Questionable Temperature Enable" on page 82
"Questionable Temperature Event Query" on page 83
"Questionable Temperature Negative Transition" on page 83
"Questionable Temperature Positive Transition" on page 84
Questionable Temperature Condition
This query returns the decimal value of the sum of the bits in the Questionable Temperature Condition
register.
The data in this register is continuously updated and reflects the current conditions.
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature:CONDition?
Example
STAT:QUES:TEMP:COND?
Preset
0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Temperature Enable
This command determines which bits in the Questionable Temperature Condition Register will set bits in
the Questionable Temperature Event register, which also sets the Temperature Summary bit (bit 4) in the
Questionable Register. The variable <integer> is the sum of the decimal values of the bits you want to
enable.
82
Sequence Analyzer User's & Programmer's Reference