6 List Sequence Measurements
Source
Marker 3
Sets the output trigger maker routing to Marker 3 for DataMarker in the whole source sequence.
Key Path
Source, List Sequencer, Trigger Type, DataMarker
Example
:SOUR:LIST:TRIG:TYPE:MARK M3
Notes
SCPI is supported after A.14.00
This an alias SCPI of :SOURce:LIST:TRIGger:OUTPut:TYPE:MARKer M3
Recommend to use :SOURce:LIST:TRIGger:OUTPut:TYPE:MARKer M3
Initial S/W Revision
A.14.00
Marker 4
Sets the output trigger maker routing to Marker 4 for DataMarker in the whole source sequence.
Key Path
Source, List Sequencer, Trigger Type, DataMarker
Example
:SOUR:LIST:TRIG:TYPE:MARK M4
Notes
SCPI is supported after A.14.00
This an alias SCPI of :SOURce:LIST:TRIGger:OUTPut:TYPE:MARKer M4
Recommend to use :SOURce:LIST:TRIGger:OUTPut:TYPE:MARKer M4
Initial S/W Revision
A.14.00
Manual Trigger Now
Pressing this key provides a software trigger event to the list sequencer. During execution of sequence, if
the sequencer is halted on any step that has been configured with a “Manual” step trigger, then this key
press will cause the sequencer to continue and execute the step.
Key Path
Source, List Sequencer
Remote Command
No remote command, front panel only.
Initial S/W Revision
A.05.00
Remote Software Trigger (Remote command Only)
During execution of a list sequence, the sequence will halt and wait at any step that has Step Trigger set to
“Bus”. Sending this command will trigger the step and continue the sequence.
Remote Command
:SOURce:LIST:TRIGger:INITiate[:IMMediate]
Example
:SOUR:LIST:TRIG:INIT
Initial S/W Revision
A.05.00
640
Sequence Analyzer User's & Programmer's Reference