6 List Sequence Measurements
Meas Setup
Key Path
Meas Setup, Basic Meas Setup, Basic Transmit Power Setup, Radio Std None
Remote Command
[:SENSe]:LSEQuencer:BTXPower:[NONE]:DIF:FILTer:ALPHa <real>
[:SENSe]:LSEQuencer:BTXPower:[NONE]:DIF:FILTer:ALPHa?
Example
LSEQ:BTXP:DIF:FILT:ALPH 0.22
LSEQ:BTXP:DIF:FILT:ALPH?
Dependencies
None.
Couplings
This parameter is grayed out when the filter type for None Radio Standard is set to flattop.
Preset
0.22
State Saved
Saved in instrument state.
Min
0.01
Max
1.0
Initial S/W Revision
A.05.00
Help Map ID
35159
Basic Frequency and Phase Error Setup
To allow the phase to be obtained, the freq offset is removed first. This then allows the phase to be
returned. In the case of multiple samples, the result returned is the average. The result returned are
Frequency Error and Phase Error with reference to the frequency set for the acquisition.
Undetermined phase changes occur between acquisitions that change frequency or instrument gain.
These cannot be calibrated out; therefore, phase error results need to be obtained from the same
acquisition.
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Basic Meas Setup
Mode
Sequence Analyzer
Initial S/W Revision
A.05.00
Help Map ID
35189
Digital IF BW
Activates the Digital IF BW function, which enables you to manually set the information bandwidth of the
analyzer. This is used to set the digital IF filter of the EXT Analyzer.
parameter_
table_
23.52941
52.94118
Key Path
Meas Setup, Basic Meas Setup, Freq and Phase Error Setup
Remote
Command
[:SENSe]:LSEQuencer:BFERor:DIF:BANDwidth[:RESolution] <freq>
Sequence Analyzer User's & Programmer's Reference