6 List Sequence Measurements
Meas Setup
:LSEQ:ACQ2:SET:MPA:EPOW 20dBm
:LSEQ:ACQ2:SET:MPA:EPOW?
:LSEQ:LIST:SET:MPA:EPOW 0,20dBm,10dBm
:LSEQ:LIST:SET:MPA:EPOW?
EPOW0,–10dBm,–20dBm
EPOW
Notes
For setting this parameter using SCPI, see
"Programming Acquisitions Via SCPI" on page 376
.
Couplings
Coupled to Basic Tx Power measurement limits if this meas is selected.
Coupled to Auto Set RF Level when set to On.
Preset
0 dBm
State Saved
Saved in instrument state.
Min
–100 dBm
Max
100 dBm
Initial S/W
Revision
A.05.00
Help Map ID
35149
Basic Meas Setup
Provides access to setup menus for the four basic measurements included in Sequence Analyzer mode:
Basic Transmit Power, Basic Frequency and Phase Error, Basic Discrete PAvT, and Basic IQ Data.
parameter_table_23.52941
52.94118
Key Path
Meas Setup
Mode
Sequence Analyzer
Initial S/W Revision
A.05.00
Help Map ID
35150
Basic Transmit Power Setup
The Basic Transmit Power measurement is available for all the radio standards that can be configured for
an acquisition: None, GSM/EDGE, WCDMA/HSPA, CDMA2000, 1xEVDO,LTE-FDD, LTE-TDD, TDSCDMA .
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Basic Meas Setup
Mode
Sequence Analyzer
Initial S/W Revision
A.05.00
Help Map ID
35151
Sequence Analyzer User's & Programmer's Reference
451