6 List Sequence Measurements
Meas Setup
Hz. If the radio band is set to a valid band, this value is interpreted as a channel number.
<
integer
> - specifies the Number of Averages for the Acquisition.
<ampl> - specifies the peak expected power in dB for the Acquisition.
<
enum
> - specifies the Instrument Gain for the Acquisition.
<
time
> - specifies the Transition Time for the Acquisition.
<
time
> - specifies the duration of the Acquisition.
<
enum
> - specifies the Input Trigger Type for the Acquisition.
<
ampl
> - specifies the Input Trigger Level for the Acquisition.
<
time
> - specifies the Input Trigger Delay for the Acquisition.
<
enum
> - specifies the Acquisition Output Trigger for the Acquisition.
<relative ampl> - specifies the dB value of Instrument Gain when Instrument Gain is set to LOW. It is Electronic
Attenuation in fact. If Instrument Gain is not set to LOW, this value is ignored. This value is optional. If it is not
set, the default value is used.
<enum> - specify the step Multiport Adapter Input path selection.
<enum> - specify the step Multiport Adapter Preamplify On/Off.
<enum> - specify the Integration type of current acquisition. this is for Span extension and Range extension.
<enum> - specify the acquisition step input path selection.
"Acquisition RF Input Port " on page 434
Dependencies
The range of sub op code values is determined by the number of Acquistions you have configured.
If you attempt to remotely set or query a sub op code that is out of range, this results in an error message.
Initial S/W
Revision
A.05.00
Modified S/W
Revision
A.14.00
Help Map ID
35231
Number of Acquisitions
This parameter is used to define the number of current active Acquisitions.
parameter_table_
23.52941
52.94118
Key Path
Meas Setup, Acquisition Setup
Remote
Command
[:SENSe]:LSEQuencer:NUMBer:ACQuire <integer>
[:SENSe]:LSEQuencer:NUMBer:ACQuire?
Example
:LSEQ:NUMB:ACQ 25
Notes
If the number is increased then any new acquisitions contain default values.
Couplings
Every acquisition must have at least one analysis step. Adding a new acquisition populates one analysis
step with default parameters.
Preset
1
Sequence Analyzer User's & Programmer's Reference