6 List Sequence Measurements
Meas Setup
TRIG:LSEQ:EXT1:SLOP?
Preset
POSitive
State Saved
Saved in instrument state.
Initial S/W Revision
A.05.00
Help Map ID
35205
External 2
Sets the trigger type to External 2 (uses a signal applied to TRIGGER 2 IN connector).
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Input Trigger Setup
Mode
Sequence Analyzer
Initial S/W Revision
A.05.00
Help Map ID
35207
Trigger Level
Sets the voltage level for the External 2 trigger.
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Input Trigger Setup
Remote Command
:TRIGger:LSEQuencer:EXTernal2:LEVel <level>
:TRIGger:LSEQuencer:EXTernal2:LEVel?
Example
TRIG:LSEQ:EXT2:LEV 1.0
TRIG:LSEQ:EXT2:LEV?
Dependencies
This parameter is used for Ext2 trigger only. Disabled if anything else.
Preset
1.2 V
State Saved
Saved in instrument state.
Min
–3.5V
Max
3.5V
Initial S/W Revision
A.05.00
Help Map ID
35209
470