6 List Sequence Measurements
Meas Setup
Initial S/W Revision
A.05.00
Help Map ID
35188
Radio Standard None
Setup parameters for Basic Transmit Power with “None” selected as the Radio Standard.
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Basic Meas Setup, Basic Transmit Power
Mode
Sequence Analyzer
Initial S/W Revision
A.05.00
Help Map ID
35153
Digital IF BW
This parameter allows you to manually set the digital IF bandwidth of the receiver path for basic transimit
power when the Radio Standard of the acquisition is set to None. This is used to set the hardware of the
digital receiver.
parameter_table_
23.52941
52.94118
Key Path
Meas Setup, Basic Meas Setup, Basic Transmit Power Setup, Radio Std None
Remote Command
[:SENSe]:LSEQuencer:BTXPower:[NONE]:DIF:BANDwidth|BWIDth[:RESolution]
<freq>
[:SENSe]:LSEQuencer:BTXPower:[NONE]:DIF:BANDwidth|BWIDth[:RESolution]?
Example
LSEQ:BTXP:DIF:BAND 1 kHz
LSEQ:BTXP:DIF:BAND?
Couplings
This parameter only takes effect when the Radio Standard of acquisition is set to NONE.
Preset
25 MHz
State Saved
Saved in instrument state.
Min
1kHz
Max
40 MHz
Help Map ID
35154
Filter Type
Selects the type of bandwidth filter that is used when the Radio Standard of the acquisition is set to None.
The choices are Gaussian, Flat top, Short Nyquist or Root Raised Cosine (RRC).
454
Sequence Analyzer User's & Programmer's Reference