5 System Functions
System
Mode
All
Remote Command
:SYSTem:MODule:MODel?
Example
:SYSTem:MODule:MODel?
Notes
This query will return model number. e.g. “M9430A”
Preset
Not affected by Preset,
State Saved
No
Initial S/W Revision
E.14.14
Module Model Serial Number (Remote Command Only)
Query only. The query “:SYST: MODule:SERial?” will return model serial number of the current module.
Mode
All
Remote Command
:SYSTem:MODule:SERial?
Example
:SYSTem:MODule:SERial?
Notes
This query will return model serial number like “SN12344321”
Preset
Not affected by Preset
State Saved
No
Initial S/W Revision
E.14.14
Power On
Enables you to select how the instrument should power on. The options are: Mode and Input/Output
Defaults, User Preset and Last State.
Key Path
System
Mode
All
Remote Command
:SYSTem:PON:TYPE MODE|USER|LAST
:SYSTem:PON:TYPE?
Example
:SYST:PON:TYPE MODE
Preset
This is unaffected by a Preset but is set to Mode on a “Restore System Defaults->All”
State Saved
No
Backwards
Compatibility SCPI
:SYSTem:PON:TYPE PRESet
the “PRESet” parameter is supported for backward compatibility only and behaves the same as
MODE.
Backwards Compatibility
Notes
The
Preset Type
key in legacy analyzers has been removed, and the
Power On
toggle key has been
replaced by this 1-of-N key in the
System
menu.
Initial S/W Revision
Prior to A.02.00
Sequence Analyzer User's & Programmer's Reference