5 System Functions
System
Key Path
System, Diagnostics
Initial S/W Revision
Prior to A.10.00
All Self Test
This key invokes all the self tests defined in the Diagnostics Self Test section.
Key Path
System, Diagnostics, Self Test
Remote Command
SYSTem:TEST:WCTS:[ALL]
Example
SYST:TEST:WCTS:[ALL]
Initial S/W Revision
A.12.50
FEC Self Test
This key invokes the EXT E6607C front end control self test. When operation is complete, the generated
test summary information is appended to log file E:\Agilent\Instrument\FECTestLog.txt. This test summary
file can be retrieved from the instrument using the MMEM set of SCPI command, once you have the fully
qualified the path and file name.
If the self test fails, the following error message will be generated:
“–330, Self-test failed, see log file E:\Agilent\Instrument\FECTestLog.txt”
If the self test passes, an advisory message “FEC self-test completed successfully” is generated.
Key Path
System, Diagnostics, Self Test
Remote Command
SYSTem:TEST:WCTS:FEC
Example
SYST:TEST:WCTS:FEC
Notes
Access log with command :
MMEM:DATA? "E:\ Agilent\Instrument\FECTestLog.txt"
Initial S/W Revision
A.12.50
Show Result
This key gives you access to show results of the following self tests:
•
Source self-test results
•
E6607C embedded MPA or E6640A/E6650A RFIO self-test results
•
E6607C FEC self-test results
Key Path
System, Diagnostics, Self Test
Initial S/W Revision
A.12.50
Sequence Analyzer User's & Programmer's Reference
203