6 List Sequence Measurements
List Sequence Measurements
To facilitate the many measurement global parameters, use of the Measurement Application Modes
parameters is made. The measurement setup sections in
describe each
measurement’s parameters.
If different Info Bandwidths were set for several measurements in an Acquisition then the List Sequencer
has to choose the highest value, acquire the Acquisition data, and then re-sample later if required.
parameter_table_23.53758
52.93301
Key Path
Remote Commands
Help Map ID
0
PVT Meas Setup
"GSM/EDGE PVT Meas Setup" on page 312
"LTE-TDD PVT Meas Setup" on page 313
"TDSCDMA PVT Meas Setup" on page 315
GSM/EDGE PVT Meas Setup
The GSM/EDGE PVT measurement used in the GSM/EDGE mode is used in the Sequencer. The Meas
parameters usable within the List Sequencer measurement are listed below. For complete details of each
parameter, please see the online help associated with the measurement.
The SCPI that is used within the Sequence Analyzer mode is the same as in the GSM/EDGE mode, except
that two additional nodes are required in the Sequence Analyzer mode: one to identify the List Sequencer
measurement and one to identify the mode from which the measurement comes. For example,
[:SENSe]:Epvtime is replaced with [:SENSe]:LSEQuencer:EDGE:EPVTime: or
:CALCulate:LSEQuencer:EDGE:EPVTime:
non_parameter_table_39.50163
6.944445
9.264707
9.264707
15.6781
Command
Description
Notes
[:SENSe]:EPVTime:BANDwidth[:RESolution]
<bandwidth>
[:SENSe]:EPVTime:BANDwidth[:RESolution]?
Info BW
Maximum
is 40MHz
[:SENSe]:EPVTime:BANDwidth[:RESolution]:TYPE
FLATtop|GAUSsian
[:SENSe]:EPVTime:BANDwidth[:RESolution]:TYPE?
Filter Type
[:SENSe]:EPVTime:BSYNc:SOURce
TSEQuence|RFBurst|NONE|RFAMplitude
[:SENSe]:EPVTime:BSYNc:SOURce?
Burst Sync
:CALCulate:EPVTime:LIMit:TEST[:STATe] OFF|ON|0|1
:CALCulate:EPVTime:LIMit:TEST[:STATe]?
Limit Test
[:SENSe]:EPVTime:MASK:SELect STANdard|CUSTom
[:SENSe]:EPVTime:MASK:SELect?
Limit Mask