6 List Sequence Measurements
List Sequence Measurements
For an example of the results returned see
"Example for two Acquisitions, each containing two Analysis
e .
In addition an Acquisition’s results can be queried on its own using:
:FETCh:LSEQuencer[1]:ACQuire[1]|2|3|4..512?
:READ:LSEQuencer[1]:ACQuire[1]|2|3|4..512?
:MEASure:LSEQuencer[1]:ACQuire[1]|2|3|4..512?
This returns all result metrics for the specific acquisition “AcqNum”. Performing a READ initiates the full
Sequence and returns only the AcqNum’s results. If the Sequence is aborted manually, then results return
NANs. If the Sequence is aborted due to “Abort on Limit Fail”, then results return NANs from point of failure
and valid results prior to that. If Sequence is aborted due to “Abort on Error”, then results return NANs from
point of failure and valid results prior to that. To find out where Limit failed or error occurred use
FETC:LSEQ3?
The result structure is very like the table for FETC:LSEQ1? above, but some fields about the whole
sequence are removed.
non_parameter_table_8.227124
74.12582
Step
Result
1
Total number of real values returned for this scpi command, including this one.
2
Reserved for Keysight use.
3
Summary Integrity value for the specified acquisition.
4
Number of Analysis Steps in the specified acquisition.
5
Summary Integrity value for Analysis interval 1
6
Analysis Interval 1 Measurement Bit Map - decimal representation
7
Integrity value for 1st measurement in Analysis Interval 1
8
Number of results in first measurement in Analysis Interval 1
9
Measurement results for first measurement in Analysis Interval 1
10
Integrity value for 2nd measurement in Analysis Interval 1
11
Number of results in 2nd measurement in Analysis Interval 1
12
Measurement results for 2nd measurement in Analysis Interval 1
13
..repeat 10 to 12 for all other measurements in Analysis Interval 1
14
Summary Integrity value for Analysis Interval 2
15
Analysis Interval 2 Measurement Bit Map - decimal representation
16
Integrity value for 1st measurement in Analysis Interval 2
17
Number of results in first measurement in Analysis Interval 2
18
Measurement results for first measurement in Analysis Interval 2
19
Integrity value for 2nd measurement in Analysis Interval 2
20
Number of results in 2nd measurement in Analysis Interval 2
21
Measurement results for 2nd measurement in Analysis Interval 2
22
..repeat 19 to 21 for all other measurements in Analysis Interval 2
226
Sequence Analyzer User's & Programmer's Reference