6 List Sequence Measurements
Meas Setup
EDGE
Selects EDGE for the Radio Standard for the current Acquisition and allows selection of the EDGE band.
parameter_
table_
23.52941
52.94118
Key Path
Meas Setup, Acquisition Setup, Radio Setup, Radio Standard
Remote
Command
-To set Radio Band individually:
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:RADio:BAND NONE | PGSM | EGSM | RGSM |
DCS1800 | PCS1900 | TGSM810 | GSM450 | GSM480 | GSM700 | GSM850
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:RADio:BAND?
-To set Radio Band as an array:
[:SENSe]:LSEQuencer:LIST:SETup:RADio:BAND <enum>,<enum>,<enum>,….
[:SENSe]:LSEQuencer:LIST:SETup:RADio:BAND?
The array form of the command is used to make a different setting for each acquisition in the sequence (the
number of array values must match the number of acquisitions; otherwise a "settings conflict" error will occur).
Example
:LSEQ:ACQ2:SET:RAD:BAND NONE
:LSEQ:ACQ2:SET:RAD:BAND?
:LSEQ:LIST:SET:RAD:BAND PGSM, NONE, GSM850, PGSM
:LSEQ:LIST:SET:RAD:BAND?
Notes
For setting this parameter using SCPI, see
"Programming Acquisitions Via SCPI" on page 376
.
State Saved
Saved in instrument state.
Range
NONE|PGSM|EGSM|RGSM|DCS1800|PCS1900|TGSM810|GSM450|GSM480|GSM750|GSM850
Initial S/W
Revision
A.05.00
Help Map ID
35046
None
Selecting None for EDGE means that in the Acquisition, the frequency of the EDGE signal is determined by
the Frequency parameter in the Acqusition setup.
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Acquisition Setup, Radio Setup, Radio Standard, GSM
Initial S/W Revision
A.05.00
Help Map ID
35233
400
Sequence Analyzer User's & Programmer's Reference