6 List Sequence Measurements
Source
Example
:SOUR:PM 1.00 rad
Preset
0.1 rad
Min
0.1 rad
Max
20.0 rad
Initial S/W Revision
A.05.00
PM Rate
Allows you to set the internal phase modulation rate.
Key Path
Source, Modulation Setup, PM
Remote Command
:SOURce:PM:INTernal:FREQuency
:SOURce:PM:INTernal:FREQuency?
Example
:SOUR:PM:INT:FREQ 40.0 Hz
Preset
400.0 Hz
Min
10 Hz
Max
40 kHz
Initial S/W Revision
A.05.00
List Sequencer
Allows you access to the sub-menus for configuring the list sequencer.
List sequences allows you to enter frequencies and amplitudes at unequal intervals in nonlinear
ascending, descending or random order. Each step within the list can also include its own waveform file for
playback, step duration, trigger event and trigger output.
The complexities involved in configuring the list sequencer do not lend itself to manual configuration;
hence the manual configuration for this feature is limited. For easier configuration of the list sequencer, it is
recommended that you use either SCPI or load a tab delimited file containing the setup parameters in a
tabular form. The details of the SCPI for configuring the list sequencer can be found in
(Remote Command Only)" on page 630
.
Once the List Sequencer has been configured using the front panel, SCPI, or loading in a tab delimited file,
the sequence must be initiated using the front panel Initiate Sequence key or the corresponding SCPI
command.
Key Path
Source
Initial S/W Revision
A.05.00
596
Sequence Analyzer User's & Programmer's Reference