![Keysight E6650A EXF User Programming Manual Download Page 72](http://html1.mh-extra.com/html/keysight/e6650a-exf/e6650a-exf_user-programming-manual_1961647072.webp)
2 Programming the Test Set
STATus Subsystem
Mode
All
Remote Command
:STATus:QUEStionable:PTRansition <integer>
:STATus:QUEStionable:PTRansition?
Example
STAT:QUES:PTR 16
Temperature summary ‘questionable asserted’ will be reported to the Status Byte Register.
Preset
32767
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Register
"Questionable Calibration Condition " on page 72
"Questionable Calibration Enable " on page 72
"Questionable Calibration Event Query " on page 73
"Questionable Calibration Negative Transition " on page 73
"Questionable Calibration Positive Transition " on page 74
Questionable Calibration Condition
This query returns the decimal value of the sum of the bits in the Questionable Calibration Condition
register.
The data in this register is continuously updated and reflects the current conditions.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:CONDition?
Example
STAT:QUES:CAL:COND?
Preset
0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Enable
This command determines which bits in the Questionable Calibration Condition Register will set bits in the
Questionable Calibration Event register, which also sets the Calibration Summary bit (bit 8) in the
72
Sequence Analyzer User's & Programmer's Reference