![Keysight E6650A EXF User Programming Manual Download Page 285](http://html1.mh-extra.com/html/keysight/e6650a-exf/e6650a-exf_user-programming-manual_1961647285.webp)
6 List Sequence Measurements
List Sequence Measurements
Preset
0 Hz
Default Unit
Hz
Initial S/W
Revision
A.14.00
Trigger Delay
Example
CALC:FPOW:POW1:DEF "TriggerDelay=0.025"
Notes
The trigger delay parameter sets the time after an external trigger is detected until the measurement is
performed.
Preset
0 s
Range
0 – 1 s
Default Unit
Seconds
Initial S/W
Revision
A.14.00
Trigger Level
Example
CALC:FPOW:POW1:DEF "TriggerLevel=2"
Notes
The trigger level parameter sets the voltage value at which an external trigger is detected.
Preset
1.2 V
Range
–5 to 5 V
Default Unit
Volts
Initial S/W Revision
A.14.00
Trigger Slope
Example
CALC:FPOW:POW1:DEF "TriggerSlope=Negative"
Notes
The trigger slope parameter indicates the direction of the edge trigger voltage for detection.
Preset
Positive
Range
Positive, Negative
Initial S/W Revision
A.14.00
Sequence Analyzer User's & Programmer's Reference
285