6 List Sequence Measurements
Meas Setup
crosses the threshold (with negative slope) and then remains below the threshold for at least the holdoff
time. In either case, the trigger event is associated with the time the level was crossed.
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Input Trigger Setup, Trigger Holdoff
Remote Command
:TRIGger:LSEQuencer[:SEQuence]:HOLDoff:TYPE BELow|ABOVe
:TRIGger:LSEQuencer[:SEQuence]:HOLDoff TYPE?
Example
TRIG:LSEQ:HOLD:TYPE BELow
TRIG:LSEQ:HOLD:TYPE?
Preset
Below
State Saved
Saved in instrument state.
Initial S/W Revision
A.11.50
Help Map ID
35340
Video
Sets the trigger type to Video (that is, IF envelope triggering).
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Input Trigger Setup
Mode
Sequence Analyzer
Initial S/W Revision
A.05.00
Help Map ID
35240
Trig Slope
This parameter controls the trigger polarity for the Video trigger. It is set positive to trigger on a rising edge,
and negative to trigger on a falling edge.
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Input Trigger Setup
Remote Command
:TRIGger:LSEQuencer:VIDeo:SLOPe POSitive|NEGative
:TRIGger:LSEQuencer:VIDeo:SLOPe?
Example
TRIG:LSEQ:VID:SLOP NEG
TRIG:LSEQ:VID:SLOP?
Preset
POSitive
State Saved
Saved in instrument state.
Initial S/W Revision
A.05.00
Help Map ID
35241
468
Sequence Analyzer User's & Programmer's Reference